How is D50 measured, and can results from different labs be compared?
D50 is the median of the cumulative particle size distribution, usually measured by laser diffraction. Different instruments and different dispersion conditions can produce quite different numbers.
The main variables are the dispersant medium (water or organic solvent), the type and dose of dispersing agent, ultrasonic power and duration, and the optical model settings. For the same powder, changing the ultrasonic time alone can shift D50 by 10-20%.
That means D50 figures from different suppliers are not directly comparable unless the test conditions match. A more reliable approach is to review the full distribution curve and ask for the test method. Our COA states the test conditions so customers can verify.